  • Applied potential: 0 – 4.8V
  • Applied potential resolution: 16 bit (smallest step: 73 µV)
  • Current ranges: 1 µA, 10 µA, 100 µA, 1 mA
  • Maximum current: ±2 mA
  • Current resolution: max. 18 bit
  • Typical noise level (1 µA range): 1 nA
  • Reference electrode input current: < 100 pA
  • Modes: potentiostatic, galvanostatic
  • Cell connections: working electrode, counter electrode, reference electrode
  • Power supply and data interface: USB