Don't use photon calibrator in PDVD data#121
Conversation
|
Hi @vpec0 , |
|
I have the impression we may need a dedicated OpHitFinder module for PDVD. Not only there are many different PDs with different SPE templates, baselines, etc, we will probably need to correct for saturation effects and I don't know if it will be very straightforward to do it with the default hit finder |
|
Got it |
I think we need to figure out, how to calibrate the channels. If we run the deconvolution first, that should take care of this. If we reconstruct from the raw waveforms directly, we need a calibrator. PDSP calibrator offers mechanism to calibrate channel-by-channel. So, ideally, we would only need to ensure proper configuration. However, that mechanism relies on the convention that the channel numbering runs from 0 through the number of channels, which is now not the case for PDVD. I see two ways around it: |
|
I have just created a draft PR, #122, that modifies the current ProtoDUNE SP calibrator to be usable for our purposes, see option a) in my previous post. With this one, we could simply set |
Current default settings set OpHitFinder to use PDSP calibrator. This messes up the output of the ophit reco.
This change then allows to remove the
PhotonCalibratorServiceProtoDUNESPservice from the configuration as well.