I've found the extra bit for insulin sensitivities > 255: it's bit 6 of the previous byte. For example, here's a bolus wizard setup record in which I change from sensitivity 35 to 400 (the max on my 522 pump):
5A 0F 1C C6 11 01 11 15 21 00 06 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 23 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 78 78 0C 64 64 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 15 21 00 06 00 00 00 00 00 00 00 00 00 00 00 00 00 00 40 90 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 78 78 0C 64 64 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 44
You can see the "00 23" in the "before" section', which changes to "40 90" in the "after" section.
When that bit gets set, it breaks the decoding of the sensitivity schedule (since that bit is set in a byte that's used for the start time of that sensitivity and needs to be masked off). The bug is in history.py, line 459.