On 5/20/19 ~15:00 I started a new set replicates in the ODIn, under a new run.
- Made sure ODIn was off, with water blank tubes in place
- Started the new run in FileMaker
- Turned the ODIn on to calibrate to blank tubes
- Transferred cultures into replicates
- Loaded tubes into ODIn
I first noticed that the file was not showing in filemaker. The file "OD1_100334.txt" was created on the server however, and data was going to this location.
Today 5/21 ~12:15 I added an additional rack (with 8 tubes) to the odin. I then noticed that the server had a new file - "OD1_100334.txt.bak". Instead of starting with my 5/20 data, it starts at 5/21 11:58. This is also where the data in "OD1_100334.txt" ends.
So I'm not sure why a new .bak file was started. I didn't restart the odin, and you can see in the first read of the .bak file that the ODIn hasn't lost track of time (value of 73332464 in the milliseconds column).
I also see 0 data in the filemaker entry. At this time, I can access both of the data files and bind them together, so I can still view the current data.