# Network drive
z_path = r"\\vs19msqapp\MOSAIQ_APP\ESCAN"
# For GUI (MessagesBoxes display errors)
import clr
clr.AddReference("System.Windows.Forms")
import os
import shutil
import sys
from connect import * # Interact w/ RS
from System.Windows.Forms import * # For GUI
def create_qa_plan():
"""Create and export a DQA plan for the current beam set
Export DICOM files to "Z:\TreatmentPlans\DQA"
If export folder with the computed name already exists, delete and recreate it
Detailed example of the code that names the new QA plan:
Existing QA plans are "Test" and "Rectal Boost DQA"
New QA plan for plan "Rectal Boost" would be "Rectal Boost DQA (1)", but name length must <=16 characters, so truncate the plan name to get QA plan name "Rectal B DQA (1)"
If we create another QA plan for "Rectal Boost", it is named "Rectal B (2)"
"""
# Get current objects
try:
beam_set = get_current("BeamSet")
except:
MessageBox.Show("There is no beam set loaded. Click OK to abort the script.", "No Beam Set Loaded")
sys.exit(1)
patient = get_current("Patient")
case = get_current("Case")
plan = get_current("Plan")
# Get unique DQA plan name while limiting to 16 characters
qa_plan_name = "{} DQA".format(plan.Name[:12].strip())
qa_plan_names = [qa_plan.BeamSet.DicomPlanLabel.lower() for qa_plan in plan.VerificationPlans]
copy_num = 0
while qa_plan_name.lower() in qa_plan_names:
copy_num += 1
copy_num_str = " ({})".format(copy_num)
qa_plan_name = "{} DQA{}".format(plan.Name[:(12 - len(copy_num_str))].strip(), copy_num_str)
# Create QA plan
iso = {"x": 21.93, "y": 21.93, "z": 0} # DICOM coordinates of "center" POI of Delta4 phantom
dg = plan.GetDoseGrid().VoxelSize
beam_set.CreateQAPlan(PhantomName="Delta4 Phantom", PhantomId="Delta4_2mm", QAPlanName=qa_plan_name, IsoCenter=iso, DoseGrid=dg, CouchAngle=0, ComputeDoseWhenPlanIsCreated=True)
# Set colorwash to CRMC standard (adapted for smaller dose)
case.CaseSettings.DoseColorMap.ColorTable = get_current("PatientDB").LoadTemplateColorMap(templateName="CRMC Standard Dose Colorwash").ColorMap.ColorTable
# Ensure phantom electronics are not excessively irradiated
qa_plan = list(plan.VerificationPlans)[-1] # Latest QA plan
point = {"x": 21.93, "y": 21.93, "z": 20} # InterpolateDoseInPoint takes DICOM coordinates
dose_at_electronics = qa_plan.BeamSet.FractionDose.InterpolateDoseInPoint(Point=point, PointFrameOfReference=qa_plan.BeamSet.FrameOfReference)
if dose_at_electronics > 20:
res = MessageBox.Show("Phantom electronics may be irradiated at {} > 20 cGy. Export anyway?".format(round(dose_at_electronics, 1)), "Create QA Plan", MessageBoxButtons.YesNo)
# Exit script if user does not want to continue
if res == DialogResult.No:
sys.exit()
# Create export folder
patient.Save() # Must save before any DICOM export
pt_name = ", ".join(patient.Name.split("^")[:2]) # e.g., "Jones, Bill"
qa_folder_name = "{} {}".format(pt_name, qa_plan_name) # e.g., "Jones, Bill Prostate DQA"
qa_folder_name = r"{}\TreatmentPlans\DQA\{}".format(z_path, qa_folder_name) # Absolute path to export folder
if os.path.isdir(qa_folder_name):
shutil.rmtree(qa_folder_name)
os.mkdir(qa_folder_name)
# Export QA plan
try:
qa_plan.ScriptableQADicomExport(ExportFolderPath=qa_folder_name, QaPlanIdentity="Patient", ExportBeamSet=True, ExportBeamSetDose=True, ExportBeamSetBeamDose=True, IgnorePreConditionWarnings=False)
except SystemError as e:
res = MessageBox.Show("{}\nProceed?".format(e), "Create QA Plan", MessageBoxButtons.YesNo)
if res == DialogResult.Yes:
qa_plan.ScriptableQADicomExport(ExportFolderPath=qa_folder_name, QaPlanIdentity="Patient", ExportBeamSet=True, ExportBeamSetDose=True, ExportBeamSetBeamDose=True, IgnorePreConditionWarnings=True)